Operation process of non-contact infrared thermometer
- 2019-01-03-

Today, Wuxi very Xiang Technology Co., Ltd. takes you to understand the operation process of non-contact infrared thermometer :



B. Object distance ratio (D: S)

Refers to the ratio of the measured distance to the surface area of the measured object. When the distance between the non-contact infrared thermometer and the measured object increases, the surface area of the thermometer and the measured object is required to be larger. D: S = 12: 1

C. Observation range

Be sure to ensure that the measured object is larger than the measurement area of the machine. The diameter of the area to be measured needs to be more than 1.5 square centimeters.

D, reflectance

Most organic materials and paints or oxidized materials have an emissivity of 0.95 (it is set in the machine). Smooth or sanded metal surfaces may cause inaccurate measurement values. To compensate, you need to cover the surface with a band or black. Paint and wait to make it the same temperature as the material below before taking a temperature measurement.

E. Press the battery switch, install the battery correctly, press the switch button to start, the LCD displays the battery symbol, the temperature value, and the value retention time is about 7 seconds.

F, use

According to the size of the measured object, select a suitable distance, aim at the measured object, press the button switch, the LCD displays the temperature value and reading.

The above is the operation process of non-contact infrared thermometer