Infrared Thermometer
SCIT-1SC (600 ~ 2000 ℃) / SCIT-2SC (300 ~ 1200 ℃) / SCIT-3S7 (100-700 ℃)
SCIT-1SC-S06 (600 ~ 2000 ℃) / SCIT-2SC-S06 (300 ~ 1200 ℃)
This series of infrared thermometer is composed of SCIT-S / SC probe and SCITV08 instrument. It uses laser to aim at the target. There are two types of laser aiming: 1) laser point aiming type; 2) laser ring aiming type.
The company also professionally supplies far-infrared thermometer products, welcome your inquiries.
1. Technical index
1) Technical index of SCIT-SC center laser spot aiming type
Medium temperature type | High temperature type | |
Type | SCIT-2SC | SCIT-1SC |
SCIT-2SC-S06 | SCIT-1SC-S06 | |
Range | 300 ~ 1200 ℃ | 600 ~ 2000 ℃ |
wavelength | 1.55μm | 0.96μm |
Distance coefficient | 100: 1 | |
Focus position | SC type: 10.4mm@1.0m (the smallest measurable target size at 1.0m is 10.4mm); SC-S06 type: 6mm @ 600mm | |
Aiming and focusing | Center laser point aiming, no focus | |
Temperature measurement accuracy | Absolute accuracy: ± 1%; repeat accuracy: ± 2 ‰ | |
Temperature measurement resolution | 1 ℃ | |
Response time | 100ms | |
Emissivity | 0.10 ~ 1.10 (adjustment step size is 0.01) | |
Temperature measurement method | Real-time, maximum, and average measurements | |
Calculation interval | Maximum / average calculation interval: 1 to 99 seconds, adjusting step size is 1 second | |
Operation and display | Five-key operation; 4-digit LED digital display, 4 to 6 LED prompts indicate alarm and temperature measurement status (can display the ring temperature of the probe) | |
The standard configuration interface | Fully isolated 14bit analog signal output: 4mA ~ 20mA Two multi-mode alarm points (AH, AL) with relay contacts and sound and light indication | |
Dimensions | Probe: Φ42mm × 158mm Control instrument type A: 160mm × 80mm × 125mm; type B: 96mm × 96mm | |
Dedicated cable | 2m (optionally up to 200m) | |
Power and power consumption | AC220V ± 10% less than 6 watts | |
working environment | Ambient temperature 0 ℃ ~ 60 ℃, humidity: 0 ~ 80% (non-condensing); Air-cooled when probe with water cooling jacket: 0 ~ 120 ℃; water-cooled: 0 ~ 175 ℃ | |
Storage temperature | 0 ℃ ~ 85 ℃ | |
Option | Fully isolated digital interface: RS232, RS485 Micro playing interface, large screen, recording unit |
The laser point emitted by the SC probe can be used to aim at the target, and the laser point represents the center of the temperature measurement area. The size of the temperature measurement area is determined according to the following figure. For example, the target diameter of the temperature measurement area at 1m is 10.4mm.
SC-S06 has a target diameter of 6mm in the temperature measurement area at 0.6m. As shown below
2) SCIT-3S7 center laser ring sighting type technical index
Type | SCIT-3S7 |
Range | 200 ℃ ~ 700 ℃ 100 ℃ ~ 500 ℃ |
Wave length | 2.30μm |
Distance coefficient | 70: 1 |
Focusing range | Focusable, 0.30m away |
Aiming method | Center laser ring aiming |
Smallest measurable Target size | 4.2mm (70: 1, at 0.30m) |
Temperature measurement accuracy | Absolute accuracy: ± 1%; repeat accuracy: ± 2 ‰ |
Temperature measurement resolution | 1 ℃ |
Response time | 100ms |
Dimensions | Φ42mm × 155mm |
Laser ring aiming: The center blank round hole is the real temperature measurement area; users can use the laser ring center round hole to aim at the measured target. Even if the red area of the laser ring is not uniform or irregular, its edges are defective, and the round hole is not in the center. It will not affect the aiming and temperature measurement. As long as there is a circular space in the laser spot (the round hole is complete), it can be aimed normally. As shown below.
Focusing: When the distance between the probe and the target is 0.25 to 0.5 meters, pull the objective lens to the outermost edge; when the distance is 1.5 meters away, push the objective lens to the innermost edge. When the distance is 0.5 to 1.5 meters, adjust the objective lens to the middle position. When focusing, the edge of the laser annulus (temperature measurement area) is clear.
2. Installation drawing
Mounting bracket (standard configuration is bent plate bracket)
3.cable connection
For details, please refer to the instruction manual.